1. Utangulizi

The Extech FL700 Fluoride Meter is a precision instrument designed for the rapid and accurate measurement of fluoride ions in various aqueous samples, including drinking water. This device integrates a sensing electrode, measurement electronics, and a digital display into a single, portable unit, making it suitable for both field and laboratory applications.

The FL700 provides direct readings of total fluoride concentrations ranging from 9.99 ppm or mg/L down to 0.1 ppm. It features automatic temperature compensation to ensure measurement accuracy across varying environmental conditions.

2. Taarifa za Usalama

Please read and understand all safety precautions before operating the Extech FL700 Fluoride Meter. Failure to follow these instructions may result in injury or damage to the instrument.

  • Always handle the fluoride electrode with care. Avoid touching the sensing element directly.
  • When working with TISAB reagent tablets or any chemical solutions, wear appropriate personal protective equipment, including gloves and eye protection.
  • Do not immerse the entire meter in liquid. The meter is designed for electrode immersion only.
  • Store the meter and reagents in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Weka mbali na watoto.

3. Bidhaa Imeishaview

The Extech FL700 Fluoride Meter is a compact and user-friendly device. Key components and features are outlined below:

Extech FL700 Fluoride Meter front view
Mbele view of the Extech FL700 Fluoride Meter, showing the display and control buttons.
Hand holding Extech FL700 Fluoride Meter
The Extech FL700 Fluoride Meter held in hand, demonstrating its ergonomic design and clear display.
Close-up of Extech FL700 electrode tip
Karibu-up view of the fluoride electrode tip, which is integrated into the meter.
Extech FL700 Fluoride Meter in use with sample
The Extech FL700 Fluoride Meter being used to measure a liquid sample, showing the meter in its stand.
Extech FL700 Fluoride Meter and included accessories
All components included with the Extech FL700: the meter, sensor cap, TISAB reagent tablets, and neckstrap.

Sifa Muhimu:

  • Urekebishaji wa pointi mbili: Ensures accurate measurements across the fluoride range.
  • Uhifadhi wa kumbukumbu: Stores up to 25 labeled readings for later review.
  • Field or lab testing: Designed for versatile use in various environments.
  • Fast response: Provides fluoride measurements in less than 1 minute.
  • Large dual LCD: Simultaneously displays ppm (or mg/L) and temperature with a bargraph.

4. Kuweka

4.1. Ufungaji wa Betri

The Extech FL700 requires four 3V button batteries (included). To install or replace batteries:

  1. Twist off the battery compartment cap located at the top of the meter.
  2. Insert the four 3V button batteries, ensuring correct polarity as indicated inside the compartment.
  3. Replace the battery compartment cap and twist to secure.

4.2. Maandalizi ya Awali

Before first use or after prolonged storage, prepare the electrode:

  1. Remove the sensor cap from the electrode tip.
  2. Rinse the electrode tip with distilled water.
  3. Condition the electrode by immersing it in a fluoride standard solution (e.g., 1 ppm or 10 ppm) for approximately 15-30 minutes.

5. Maagizo ya Uendeshaji

5.1. Kuwasha/Kuzima

Bonyeza kwa WASHA/ZIMWA kitufe cha kuwasha au kuzima mita.

5.2. Calibration (Two-Point)

Accurate calibration is crucial for reliable measurements. The FL700 supports two-point calibration.

  1. Prepare two fluoride standard solutions, typically one low concentration (e.g., 1 ppm) and one high concentration (e.g., 10 ppm). Ensure the solutions are at a stable temperature.
  2. Suuza electrode na maji distilled na msamehevu kavu.
  3. Immerse the electrode into the first standard solution.
  4. Bonyeza na ushikilie CAL / KUMBUKA button until 'CAL' appears on the display. The meter will automatically recognize the standard and calibrate. Wait for the reading to stabilize.
  5. Rinse the electrode again and immerse it into the second standard solution.
  6. Bonyeza na ushikilie CAL / KUMBUKA button again. The meter will complete the two-point calibration.
  7. The meter is now calibrated and ready for measurement.

5.3. Kuchukua Kipimo

To measure fluoride concentration in a sample:

  1. Hakikisha mita imepimwa.
  2. Suuza electrode na maji distilled na msamehevu kavu.
  3. Add TISAB (Total Ionic Strength Adjustment Buffer) reagent to your sample according to the reagent instructions. TISAB ensures that the fluoride ions are in a measurable form and minimizes interference.
  4. Immerse the electrode tip into the prepared sample solution. Ensure the tip is fully submerged.
  5. Allow the reading to stabilize on the display. The large dual LCD will show the fluoride concentration in ppm (or mg/L) and the temperature.
  6. Ili kushikilia usomaji wa sasa, bonyeza MODE / KUSHIKILIWA kitufe. Bonyeza tena ili kuachilia.

5.4. Recalling Stored Readings

The meter can store up to 25 labeled readings.

  • To recall stored readings, press the CAL / KUMBUKA button briefly when not in calibration mode.
  • Tumia MODE / KUSHIKILIWA button to scroll through the stored readings.

6. Matengenezo

6.1. Kusafisha kwa Electrode

Regular cleaning of the electrode ensures optimal performance.

  • Baada ya kila matumizi, suuza elektrodi vizuri na maji yaliyosafishwa.
  • If the electrode becomes fouled or readings become erratic, clean it with a mild detergent solution or a specialized electrode cleaning solution. Rinse thoroughly with distilled water afterward.

6.2. Hifadhi

Proper storage extends the life of the electrode and meter.

  • Always replace the sensor cap on the electrode tip when not in use. The cap should contain a small amount of electrode storage solution or a fluoride standard solution to keep the membrane hydrated. Do not store dry.
  • Hifadhi mita mahali pa baridi, kavu.

6.3. Kubadilisha Betri

Replace batteries when the low battery indicator appears on the display or when the meter fails to power on. Refer to Section 4.1 for battery installation instructions.

7. Utatuzi wa shida

TatizoSababu inayowezekanaSuluhisho
Kipima hakiwakiBetri zilizokufa au zilizosakinishwa vibayaAngalia polarity ya betri; kuchukua nafasi ya betri.
Usomaji usio sahihi au usio thabitiDirty electrode; air bubbles on electrode; insufficient sample volume; expired or incorrect TISAB.Clean electrode; gently tap electrode to remove bubbles; ensure electrode is fully submerged; check TISAB.
Meter fails to calibrateIncorrect standard solutions; dirty electrode; electrode not conditioned; damaged electrode.Verify standard concentrations; clean and condition electrode; replace electrode if damaged.
Kiashiria cha chini cha betriBetri ziko chini.Replace all four 3V button batteries.

8. Vipimo

VipimoMaelezo
MtengenezajiExtech
Nambari ya SehemuFL700
Uzito wa Kipengee7.7 wakia
Vipimo vya BidhaaInchi 6.8 x 1.6 x 1.4
Nambari ya Mfano wa KipengeeFL700
BetriBetri 4 za Lithium Metal zinahitajika (zimejumuishwa)
MtindoFL700
Kiasi cha Kifurushi cha Bidhaa1
Idadi ya Vipande1
Vipengee vilivyojumuishwaKitengo, Mwongozo wa Maelekezo
Aina ya Kiini cha BetriAlkali
Tarehe ya Kwanza InapatikanaJulai 15, 2008

9. Taarifa za Udhamini

The Extech FL700 Fluoride Meter comes with a full one-year warranty on parts and workmanship, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modification.

10. Msaada

For technical support, service, or additional information, please visit the official Extech webtovuti au wasiliana na idara yao ya huduma kwa wateja. Unaweza pia kupata habari zaidi na bidhaa kwenye Extech Store on Amazon.

Nyaraka Zinazohusiana - FL700

Kablaview EXTECH 45170 4-in-1 Humidity, Joto, Airflow, na Mwongozo wa Mtumiaji wa Mita Mwanga
Mwongozo wa mtumiaji wa EXTECH 45170, mita 4-katika-1 kupima unyevu, halijoto, mtiririko wa hewa na mwanga. Inajumuisha uendeshaji, vipimo, na maelezo ya udhamini.
Kablaview Manuale Utente Misuratore di Radiazione UV-AB Extech UV505
Manuale utente completo per il Misuratore di Radiazione UV-AB Extech UV505. Jumuisha uendeshaji wa istruzioni, mbinu maalum, taarifa za sicurezza na manutenzione.
Kablaview Mwongozo wa Mtumiaji wa Mita Mwanga ya UV505 UV-AB
Mwongozo wa mtumiaji wa Extech UV505 UV-AB Light Meter. Hati hii inatoa maelezo ya kina kuhusu vipengele vya kifaa, uendeshaji, vipimo, tahadhari za usalama na matengenezo ya kupima mwanga wa UVA na UVB.
Kablaview Extech EX810 1000 Amp Klemmeter ilikutana na Kipima joto cha IR Gebruikershandleiding
Gedetailleerde gebruikershandleiding voor de Extech EX810 1000 Amp klemmeter alikutana ingebouwde IR kipimajoto. Maagizo ya Bevat, veiligheidsrichtlijnen en technische specificaties voor nauwkeurige metingen van spanning, stroom, weerstand, capaciteit, frequentie en temperatuur.
Kablaview Mwongozo wa Mtumiaji wa Msururu wa Extech EX650: 600A Kweli RMS Digital Clamp Mita
Mwongozo wa kina wa watumiaji wa Msururu wa Extech EX650 wa 600A True RMS digital clamp mita, mifano ya kufunika EX650 na EX655. Pata maelezo kuhusu utendakazi, vipengele, usalama na vipimo vya ujazo sahihi wa AC/DCtage, sasa, upinzani, na vipimo zaidi.
Kablaview Mwongozo wa Mtumiaji wa Extech DM100 Pocket Multimeter
Mwongozo wa mtumiaji wa Kipima-Kipimo cha Mifuko cha Extech DM100 4000 Count Autoranging Pocket, kinachohusu usalama, vipimo, uendeshaji, vipengele, na matengenezo.